Sterilizer common malfunctions and Alarm list

no. | trouble | cause | The elimination method |
1 | The cabinet body water remained | Vacuum pump, vacuum valve switch out of sequence | Operate according to the use of sterilizer operating guide manual |
The air return inlet water disk block | Dredge | ||
The water in the steam water separator is not eliminated | Drainage | ||
during the Humidifying water entered | inspection The steam pressure of steam generator, controlling the wetting velocity | ||
condensed water too much residual in the Steam filter | Open the steam filter sewage outlet discharge water | ||
2 | Cabinet air leakage | Door strip aging, damage | Replace door strip |
Door seal groove debris | removal of debris | ||
The failure of safety valve | Send inspection institute for maintenance | ||
relevant valve,pipe leakage | fasten again or replace | ||
3 | Vacuum pump does not work | Circuit breaker trip | Find out the cause, repair |
Thermal relay protection | Pump working current is too large, should stop work, find out the reason, overhaul | ||
4 | Vacuum pump Water Leakage | water Seal sealing ring damage | replace |
Flange loose | fasten | ||
5 | Vacuum pump abnormal sound | lack of water or the water is too small | Check the vacuum pump water supply |
the vacuum valve not Open | Open the vacuum valve | ||
There is debris in impeller cavity | Remove debris | ||
6 | The Water pump Leakage | The sealing ring is damaged | Replace sealing ring |
7 | water Pump do not work | Circuit breaker trip | Find out the cause, repair |
Thermal relay protection | Pump working current is too large, should stop work, find out the reason, overhaul | ||
8 | Vacuum pumping speed too slow | vacuum pump Lack of water or water is too small | Check the vacuum pump water supply |
The cabinet is provided with leakage | Locate the leak point, repair | ||
making vacuum system leakage | Find vacuum pipeline leak points, fastening | ||
9 | EO cannot be added | eo feeding pipeline plugging | Dredge |
The ambient temperature is too low | EO cylinder insulation | ||
10 | Temperature raise is too slow | circuit break of electric heating tube | To repair the circuit of electric heating tube |
Water tank temperature is too low | According to the instructions for use in operational guidelines re-setting tank temperature value | ||
11 | Temperature, humidity, pressure display inaccurate or out of control | Sensor damage | change |
Sensor supply voltage is insufficient | Find out the reason, replace | ||
12 | IPC cannot normally start or work | Fault of industrial control computer system | Looking for industrial computer manufacturer or IPC professional repair personnel to repair |
Viral infection | Use antivirus software kill the virus | ||
Memory bar loose | Open the case to insert memory bar again | ||
13 |
The door cannot be opened |
The air pressure of air source is not enough | check air pipe pressure |
Throttle valve failure | adjust or change | ||
Sterilized goods expansion extrude door | wait the products cooled and shrink | ||
Cylinder solenoid valve failure | replace | ||
Mechanical obstruction |
Check the door moving parts is flexible, or effectively adjust or repair | ||
14 | The door cannot be closed | The air pressure of air source is not enough | check air pipe pressure |
Throttle valve failure | adjust or change | ||
Cylinder solenoid valve failure | replace | ||
Mechanical obstruction | Check the door moving parts is flexible, or effectively adjust or repair | ||
15 | Door seal can not inflate | Door sealing electromagnetic valve is damaged | Repair or replace |
Position switch failure | Adjust or replace | ||
The door is not closed in place | Check the obstacles | ||
16 | Door cylinder leakage | The damage of the sealing ring | replace |
Cylinder joint loosening | fasten | ||
17 | Internal circulation system does not work | Fault of magnetic transmission device | Check the magnetic transmission device |
The pipeline is not smooth | clean the pipe |